Sale!

Dräger 1/A Detection Tube for Simultaneous Hydrochloric Acid and Nitric Acid Vapor Monitoring – Part No. 8101681

Original price was: $152.00.Current price is: $76.00.

In stock

Quantity
Category: Brand:

Description

Dräger-Tube Hydrochloric Acid/Nitric Acid 1/a (10)
PN: 8101681

Dräger-Tubes™ are used together with a Dräger tube pump for measurements of gases, vapors and aerosols, usually in air, and are one of the classic methods of gas analysis. This Dräger-Tube is used for the detection of hydrochloric acid/nitric acid bromide and has the following standard measuring ranges: Hydrochloric acid 1-10 ppm Nitric acid 1-5 ppm Measurements: 10


Application Range
Standard Measuring Range 1 to 10 ppm / 1 to 15 ppm
Number of Strokes n 10                / 20
Time for Measurement approx. 1.5 min  / approx. 3 min
Standard Deviation ± 30%
Color Change blue to yellow
Ambient Operating Conditions
Temperature 5 to 40 °C for HCl
For HNO3 measurements, the tube scales are applicable only to 20 °C. Where the temperature is different,  the measuring result must be multiplied as follows:
Temperature            Factor 
40                            0.3
30                            0.4
10                            2
Absolute Humidity max 15 mg H2O/L 
 Cross Sensitivity
50 ppm nitrogen dioxide result in almost the same reading as 2 ppm nitric acid. 10 ppm hydrogen sulfide or 5 ppm nitrogen dioxide do not affect the reading. Chlorine concentrations in excess of 1 ppm change the entire indicating layer to yellow-green